引用本文
  • 许楠,潘海红,刘晓薇,等.第2代双源CT冠状动脉个性化扫描方案效果评价[J].同济大学学报(医学版),2014,35(1):76-79.    [点击复制]
  • XU Nan,PAN Hai-hong,LIU Xiao-wei,et al.Different scan modes of 2nd-generation dual-source CT for CT coronary angiography[J].同济大学学报(医学版),2014,35(1):76-79.   [点击复制]
【打印本页】 【在线阅读全文】【下载PDF全文】 查看/发表评论下载PDF阅读器关闭

←前一篇|后一篇→

过刊浏览    高级检索

本文已被:浏览 269次   下载 279 本文二维码信息
码上扫一扫!
第2代双源CT冠状动脉个性化扫描方案效果评价
许楠,潘海红,刘晓薇,林翌,乔琛,张蕾
0
(同济大学附属东方医院医学影像科,上海200120)
摘要:
目的探讨第2代双源CT的三种扫描模式在不同心率、心律患者冠状动脉检查的应用效果,比较其图像 质量和扫描剂量。方法根据不同心率、心律将166例冠状动脉检查者分为A、B、C组(A组:平稳低心率组,B 组:平稳高心率组,C组:不稳定心律组),分别行炫速双源CT Flash、Sequence或Routine扫描模式检查,记录扫描 时的心率、心率变异幅度,和扫描剂量,并评价冠脉三支主干的图像质量(0~3分)。结果Flash法扫描患者67 例,平均扫描剂量为(0.73±0.09)mSv;Sequence法扫描35例,平均扫描剂量为(3.73±1.27)mSv;Routine法扫描 64例,平均扫描剂量为(9.07±2.96)mSv,三种扫描模扫描剂量差异有统计学意义(P<0.01)。三种扫描模式下, 各支冠状动脉主干的图像质量差异无统计学意义(P>0.05)。心率和图像质量有相关性(P<0.01),而心率变异 幅度与前降支的图像质量有相关性(P<0.01),与回旋支和右冠状动脉的图像质量无相关性(P>0.05)。结论第2代双源CT能依据患者的心率、心律情况进行个性化选择冠脉CT扫描模式,在保证较高的图像质量的同时可 以最大程度降低患者的扫描剂量。
关键词:  冠状动脉  体层摄影术,X线计算机  低辐射剂量
DOI:10.3969/j.issnl008-0392.2014.01.016
基金项目:
Different scan modes of 2nd-generation dual-source CT for CT coronary angiography
XU Nan,PAN Hai-hong,LIU Xiao-wei,LIN Yi,QIAO Chen,ZHANG Lei
(Dept. of Radiology, East Hospital, Tongji University, Shanghai 200120, China)
Abstract:
Objective To evaluate the application of three scan modes for CT coronary angiography by using 2nd-generation dual source CT. Methods One hundred and sixty six patients underwent CT coronary angiography with 2nd-generation dual source CT scanner (DSCT:SOMATOM Definition Flash),among them 67 patients (HR<70 bpm with regular rhythm) were examined with flash scan mode, 35 patients (HR≥70 bpm with regular rhythm) with sequence scan mode, and 64 patients (irregular rhythm) with conventional spiral mode, respectively. The radiation dose and image quality of three main coronary artery branches of these scan modes were compared. Results The average doses used for flash,sequence and conventional scan modes were (0.73±0.09), (3.73±1.27) and (9.07±2.96) mSv, respectively (P<0.01). There were no significant differences in the image quality of three main branches of coronary arteries among 3 different scan inodes (P>0.05). The image quality was significantly correlated with heart rate (P<0.01). The variation of heart rate was correlated with the image quality of LAD (P<0.01),but not correlated with that of LCX and RCA (P>0.05). Conclusion The scan mode of 2 nd-generation DSCT image can be chosen according to heart rate and rhythm of patients to achieve satisfactory imaging quality and low radiation dose.
Key words:  coronary artery  tomography,X-ray computed  low radiation dose

您是第5128390位访问者
版权所有《同济大学学报(医学版)》编辑部
主管单位:教育部 主办单位:同济大学
地  址: 上海四平路1239号 邮编:200092 电话:021-65980705 E-mail: yxxb@tongji.edu.cn
本系统由北京勤云科技发展有限公司设计